Electron Microscopy
Electron microscopy (EM or TEM) is a method of examining tissue specimens using an electron beam instead of ordinary light. It is useful in certain conditions where ordinary light microscopy or other laboratory tests are incapable of establishing a diagnosis. Tissue must be specially handled and processed for this test, including the use of special fixatives (glutaraldehyde and paraformaldehyde) and special histologic techniques (embedding in plastic, sectioning with special microtomes and special types of stains such as osmium tetroxide). Tissue cannot be handled using routine procedures for light microscopy or else the quality will suffer and a diagnosis may not be possible. The results of EM examination produce a black-and-white image and usually photographs for archival purposes. With some specimens such as neoplasms, the results of EM are correlated with the routine light microscopic appearance and immunohistochemistry profile to arrive at a final diagnosis. Electron microscopy is a useful tool and provides information not available by other means. It is routinely used in the evaluation of primary medical renal disease of unknown etiology, evaluation of poorly differentiated malignancies, amelanotic melanoma, neuroendocrine neoplasms, mesothelioma, Langerhans cell histiocytosis, some leukemias and lymphomas such as acute promegakaryocytic leukemia, evaluation of muscle biopsies in certain conditions such as mitochondrial myopathies, evaluation of certain viral diseases such as herpes virus, evaluation of immotile cilia syndrome from nasal or bronchial biopsies, certain liver conditions such as Dubin-Johnson syndrome and evaluation of certain heart disorders such as suspected adriamycin toxicity. EM is routinely used for the examination of kidney biopsies as part of the workup of abnormal renal function in some patients.
In the modern era of immunohistochemistry light micrscopy, electron microscopy is used less often for the diagnosis of poorly differentiated malignancies than it once was. It is still a necessary and valuable tool for the evaluation of non-tumor related kidney biopsies.
Findings vary and are reported depending on the type of specimen and clinical question. For example, the finding of pre-melanosomes in a poorly differentiated malignancy is indicative of malignant melanoma and the absence of dynein arms in a nasal ciliary biopsy is indicative of immotile cilia syndrome.
Specimens should be submitted fresh and cut into small pieces (approximately 1mm) before placing in glutaraldehyde or paraformaldehyde. Specimens should not be placed in any other type of fixative or results will be sub-optimal. Formalin fixed paraffin embedded tissue is not recommended.
- Ultrastruct Pathol. 1999 Mar-Apr;23(2):63-77. [PMID: 10369101]
- J Neuropathol Exp Neurol. 2009 Aug;68(8):833-44. [PMID: 19606069]
- Appl Immunohistochem Molecul Morphol. 2000; 8(2): 87-97. [PMID: 10937055]
- Guides to Clinical Aspiration Biopsy, Diagnostic Immunocytochemistry and Electron Microscopy. New York, NY: Igaku-Shoin (1991), Pg. 11-25.